Who's Online
There currently are 5813 guests and 1 member online.
|
Categories
|
Information
|
Featured Product
|
|
|
|
|
There are currently no product reviews.
;
Very clear scan, I recommend it. Definitely a must have for any 3362 owner.
Alpine could have written a slightly more complete manual, though. It's already pretty huge, but the unit has so many functions, I feel some more explanation would have been better.
Yamaha's manual of their comparable YDSP-1 is a little better in my opinion.
;
Immediate response with excellent service - highly reccommended
;
Correct manual received and of good quality but the contents of the file for the Service Manual for the same monitor is for a 20" TV not the RGB Monitor.
;
The owners manual for the Pioneer VSX-D603S was exactly what I needed. I received it in a timely manor and it was priced fairly. I definately will use your servise again. Thank You
;
This is a great manual. It was easy to read and very accurate. The size of the manual is small so email is no problem. I highly recommend it. The manual is very hard to find other than this website.
Model 4342A
Section Figure
TTT
3-6
Mechanical
zero
adiustment
The meter is properly zero-set when the pointer sets exactly over the zero calibration scale mark and the instrument is in its normal operating environment. To check the meter mechanical zero, turn the instrument off and allow 30 seconds to completely deenergizethe z strument. To obtain maximum accuracy and mechanical stability, if the meter is not over zero, zero-set the meter as follows: a. b. Rotate meter pointer adjustment screw @ clockwise is moving toward zero in an upscale direction. Continue rotating screw clockwise and stop exactly at zero. If the pointer overshoots, the adjustment screw clockwise to do steps until meter
when pointer is continue rotating a and b once again.
C.
When the pointer is exactly over zero, rotate adjustment screw slightly counterclockwise to relieve tension on pointer suspension. If pointer moves off zero, repeat steps a, b and c, but rotate less counterclockwise. zero adiustment
Electrical
The meter pointer should set exactly over the zero scale mark when instrument is turned on and nothing is connected to measurement terminals. Turn the instrument on and allow at least 15 minuts warm-up time to let the instrument reach a stable operating condition. If meter pointer is not over zero, zero-set the meter as follows: a. b. Set FREQUENCY RANGE selector 001) and Q RANGE@ to 1000. Adjust pointer rear panel is exactly @to 22k - 70k (10k - 32k for so that the meter Option
METER ZERO ADJ control@ over zero.
Figure
3-6.
Zeroing
Procedure
(sheet
2 of 2).
3-9
|
|
|
> |
|